TP-U is a body surface temperature measuring unit that installs 2000 pixels thermal imaging sensor. It quickly measures the body surface temperature and if detects the alarm point tenperature (ex. 38°C) then the failure lamp lightens. Install at employee entrance, school entrance, croded facilities, etc and simply check the fever..
| Measures body surface temperature just in 3 seconds. |
| Easy operation and self check by confirming standby, good and failure lamps. |
| Hygienic because non-contact measuring. |
| Prepared for stand-alone operation. |
| Exclusive application software. |
| Model | Specification |
| TP-U0260ES | 20 to 40°C, Viewable Angle: 60°x60°, Ethernet |
| SPN(Accessory) | Measuring hood |
| Temp Range: | 20 to 40°C |
| Accuracy rating: | ±0.5°C at 37°C (Ambient temperature 25°C) |
| Detector: | Thermopile array Approx. 2000 Pixel |
| Wavelength: | Center wavelength 10μm |
| Viewable Angle: | 60°×60° |
| Radius Resolution: | 21.8mrad |
| Frame Speed: | 1Hz |
| Focus: | Fixed focus |
| Communications Interface: | Ethernet---10BASE-T/100BASE-TX |
| Contact Output: | 2 points(Non voltage contact output) |
| Power Supply: | 100 to 240V AC(universal power supply) |
| Working Temp: | 0 to 40°C |
| Working RH: | 10 to 80%RH (No dew condensation) |
| Casing: | Polycarbonate (Color: Black) |
| Weight: | Approx. 700g |
| Warm up time: | Approx.60min |
| Standards: | CE (EN61326 Annex A) |
| Temp Judgment Function: | Set per 0.5°C |
| Designate Alarm Judgment Area: | Change Judgment Area |
| Comment Function: |
Display comments to the judgment result of value exceeding the setting. |
| Data Storage: |
Auto save by measurement data (CSV) and thermal image (JPEG) |
| Measurement Temp. Correction: |
Manual Correct: Set per 0.1°C by PC. Auto Correct: The unit automatically set the correction value. |
| Language: | Japanese, English, Chinese and Korean |
Entrance check
where the ambient temperature is stableReal-time trend screen
The temperature inside of the mouth is resistible to the outside temperature and stable measurement is possible. (Recommended Function: Manual Correct)

Entrance check
where the ambient temperature fluctuates.
When using semi-outdoor, Avoid direct sunlight and in the weather.
(Recommended Function: Auto Correct)

1. Confirm the orange lamp is lighting.
Confirm that the unit is stand-by. Body surface temperature might change by the ambient temperature.

2. Measure body surface temperature.
Lean your head to the sensor about 15cm.
Wait 3 seconds until it sounds "Be-Be-Beep"

COnfirm the judgment result.
Green lamp lightens when below the set alarm point (ex. 38°C) or red lamp blinks.

Reconfirmation