The HN-C series is reliable and accurate temperature/ humidity meter with the measuring humidity range of 0 to 100%RH by using a new developed humidity sensor.
Models with temperature output of 3-wire type Pt100 is available.Wall-mounting type, separated type and insertion type are prepared for various applications.
Heat-resistive separated type is available for using in the environment of -40 to 180°C.
| Improved repeatability and reliability by a new developed humidity sensor |
| Measurement of relative humidity 0 to 100%RH by capacitance type humidity sensor |
| Three types of wall-mounting type, separated type, heat-resistive separated type and insertion type (soft type and hard type) are prepared for various applications. |
| Heat-resistive separated type can be used in the environment of -40 to 180°C |
| Easy maintenance by complete interchangeable sensors |
| DIN rail installation enabled (except insertion type) |
| CE conformance |
| Models | Types | Temperature output signal |
| HN-CQA | Wall-mount type | 4 to 20mADC (2-wire transmission) |
| HN-CVA | Insertion type (soft type) | |
| HN-CPA | Insertion type (hard type) | |
| HN-CFA 〈 | Separated type | |
| HN-CGA 〈 | Separated heat-resistive type | |
| HN-CQB | Wall-mount type | Pt100 3-wire system |
| HN-CVB | Insertion type (soft type) | |
| HN-CPB | Insertion type (hard type) |
| Type | Wall-mount type | Separated type |
Separated heat resistive type |
Insertion type (soft type) |
Insertion type (hard type) |
|
| Model | HN-CQ😠| HN-CFA | HN-CGA | HN-CV😠| HN-CP😠| |
| Humidity | Sensor | High Polymer capacitance type | ||||
| Measuring range | 0 to 100%RH (no dew condensation) | |||||
| Accuracy rating | ±2%RH (at 0 to 90%RH and 25°C), ±3%RH (at 90 to 95%RH and 25°C) | |||||
| Relative humidity/temperature coefficient | ±0.1%RH/°C (at the range of 5 to 80°C) | |||||
| Response time | Within 15 seconds (at 90% response, 25°C and under 0.1m/s wind flow) | Within 1 minute (at the same condition shown left) | ||||
| Output signal | 4 to 20mADC, 2-wire transmission (against 0 to 100%RH) | |||||
| Load resistance | 500Ω or lower (at 24VDC power supply) | |||||
| Temperature | Sensor | Platinum resistance thermometer Pt100 JIS class A | ||||
| Measuring range | 0 to 50°C | -40 to 80°C | -40 to 180°C | 0 to 50°C | 0 to 80°C | |
|
Accuracy rating (for 2-wire transmission) |
±0.3°C | ±0.5°C | ±0.3°C | |||
| Output signal |
4 to 20mADC, 2-wire transmission (against measuring range) for HN-CðŸ˜A Pt100, 3-wire system for HN-CðŸ˜B |
|||||
|
Load resistance (2-wire transmission) |
500Ω or lower (at 24VDC power supply) | |||||
| Sensor interchangeability | Plug-in interchangeable | Plug-in interchangeable (including cable) | Plug-in interchangeable | |||
| Working temperature range | Meter | 0 to 50°C | ||||
| Sensor | 0 to 50°C | -40 to 80°C | -40 to 180°C | 0 to 50°C | 0 to 80°C | |
| Power Voltage | 12 to 24VDC ± 1V | |||||
| Connection | Crump type no-screw terminal connection | Terminal connection (M4 x 5) | ||||
| Material | Meter | Resin | Die-cast aluminum | |||
| Protective tube | ---- | Resin | SUS | |||
| Mounting | Screw fixing (M4 x 2) or DIN rail mounting | Screw (M4 x 2) | Flange (JIS 5K 25A) | |||
| Weight | About 160g | About 220g | About 240g | About 1.6kg | ||
| Accessory | Fixtures for wall-mounting | ---- | ---- | |||
| Others | ---- | Cable length: 1m, 2m, 3m (to be specified) | ---- | ---- | ||
| Performance Specification Sheet | Temperature/Humidity Meters | HN-C series | Download |