IR-ESF Series Fiber Optic Scanning Type Radiation thermometers

The IR-ESF series is applicable to measure width-direction temperature patterns of moving objects.
The system is consisted of a scanning unit, a display/operation unit with a built-in exclusive software for operation and graphic display, and connection cables.

Two kinds of elements, InGaAs for medium temperature and Si, and two kinds of measuring systems, single-color type and 2-color type, are available.
The small sized (100 x 150 x 85mm) and light weight (about 1.3kg) scanning unit is easy to install.
The display/operation unit with a built-in exclusive software displays thermal-images in pattern, trend and color modes.
The output is 0 to 5VDC for 7 position points + 1 peak point or 3 zone peak points.
For targeting of objects, a direct viewfinder type or laser spotting type are available. (option)

 

 

Models Type
IR-EFS Main unit Single-color Si element
IR-EFI Single-color InGaAs element
IR-EFQS 2-color Si element
IR-EFQI 2-color InGaAs element
IR-ESF 〈 〈 Scanning unit
IR-EP 〈 5F 〈 Display/operation unit For single-color
IR-EQN5F 〈 For 2-color
IR-ZFN 〈 〈 Fiber optic cable
IR-ZEFC 〈 〈 Cable between scanning unit and main unit
IR-ZE 〈 C 〈 〈 〈 Cable between display/operation unit and main unit
IR-ZERC 〈 〈 〈 Cable for output from display/operation unit

 

 

Main Unit
  IR-EF IR-EFQ
Element InGaAs or Si
Measuring system Single-color 2-color
Measuring temperature InGaAs: 300 to 750°C, 350 to 900°C
400 to 1100°C, 500 to 1300°C
Si: 600 to 1100°C, 700 to 1300°C
800 to 1600°C, 900 to 2000°C.
InGaAs: 350 to 600°C, 400 to 700°C
600 to 1100°C, 800 to 1600°C
Si: 1000 to 1600°C, 1200 to 2000°C.
  InGaAs: Lower than 1000°C: ±5°C
1000°C or higher: ±0.5% of measured value
Si: Lower than 800°C: ±4°C
800°C to lower than 1500°C: ±0.5% of measured value
1500°C to lower than 2000°C: ±1.0% of measured value
InGaAs: Lower than 1000°C: ±5°C
1000°C or higher: ±0.5% of measured value
Si: Lower than 1500°C: ±0.5% of measured value
1500°C to lower than 2000°C: ±1.0% of measured value
Fiber optic cable Standard IR-ZFN04 with metal protective tube 4m (max 20m)
Output to display/operation unit Temperature signal: 4 to 20mA
Scanning synchronous signal: Open collector
Connection to display/operation unit Exclusive connection cable
Rated power voltage 100 to 120VAC or 200 to 240VAC (to be specified), 50/60Hz
Weight About 7.2kg About 5.5kg
Scanning unit
Scanning system Optical axis scanning by a rotating mirror
Scanning angle 90°
Scanning speed About 5 times/second
Measuring diameter/
measuring distance
Φ40/4000mm or Φ20/300mm (IR-EFQ only) (The relation of the measuring diameter and the measuring distance differs depended on a fiber optic cable and a lens assembly.)
Connection to display/operation unit Fiber optics and exclusive connection cable
Power supply 24VDC (supplied from main unit)
Weight About 1.3kg
Display/operation unit
OS Windows95 or WindowsNT4.0 (to be specified) (Windows NT4.0 only for IR-EFQ)
CRT 800 x 600 dots
Screen Thermal-image display in pattern, trend and color modes
Output 0 to 5VDC (without isolation)
Output points (1) 7 position points + 1 peak point or (2) 3 zone peak points at sampling mode
Software Built-in standard software for Ffber optic scanning type radiation thermometer
Rated power voltage 100 to 240VAC 50/60Hz
Weight About 8.5kg

 

 

Instruction Manual Scanning Type Radiation thermometers IR-ESF Download
Leave a feedback